Taking Toc Measurements; The Menu Screen; Setup; Setting The Analyzer Mode - GE Sievers 500 RL Operation And Maintenance Manual

On-line total organic carbon analyzer
Hide thumbs Also See for Sievers 500 RL:
Table of Contents

Advertisement

Chapter 4: Basic Analyzer Operation
3.
The Status area displays information about the current operation mode, the Analyzer's current status
(Initializing, Idle, Analyzing, etc.), a countdown timer showing the time remaining in the current analysis,
and the repetition and rejection count, if applicable.

Taking TOC Measurements

After you have set the analysis mode and have configured the mode settings (as described in "Setting the
Analyzer Mode" on page 69) you are ready to begin taking TOC measurements. Simply press the Start Analysis
button to begin.

The Menu Screen

The Menu screen provides you with access to all Analyzer settings and configuration options. Note that if
Password protection or DataGuard is enabled, some options may be available only to certain users. See
Figure 22 on page 102 for a menu map that indicates the required User Level to access various settings.
The Menu screen contains the five main configuration categories; press a tab to display the available options:
Setup — Set the Analyzer mode and change parameters such as the number of repetitions and the
number of rejects.
Data — Display and export historical data stored in RAM.
I/O — Configure the input and output of data from the Analyzer.
Maintenance — Display information about the Analyzer's consumables levels, system information, and
error messages. Also perform calibration and system suitability operations.
DataGuard — Configure the access settings and user passwords for your Analyzer.
From any tab under the Menu screen, you can press the Start Analysis button to initiate TOC measurements.

Setup

Options on the Setup tab let you specify preferences and basic operation settings for the Analyzer configuration.
If the sample conditions for your application remain relatively constant, you should seldom need to configure
the settings here.

Setting the Analyzer Mode

The Sievers 500 RL On-Line TOC Analyzer can operate in up to four different modes, depending on your
configuration. To set the Analyzer mode, follow these steps:
GE Analytical Instruments ©2010
69 of 226
DLM 74001-04 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents