Chapter 6: Calibration and System Protocols

Required Calibration Supplies

The number of standards required for the calibration procedure depends on whether you choose a single-point
or multi-point calibration. Standards vials are inserted into the iOS System as prompted by the Analyzer's on-
screen instructions.
A single-point calibration should require approximately 2.75 hours to complete, and multi-point calibration
should require approximately 4.5 hours to complete.
Sievers Standards for all calibration and verification procedures should be purchased from GE Analytical
Instruments. Standards sets for calibration and verification can be purchased individually, or as vial sets for use
with a Super iOS System. All standards are provided in 40-mL and 30-mL vials. Standards sets for validation
(System Suitability, Accuracy & Precision, Linearity, Robustness, Specificity, and JP Protocol) can be purchased
individually, or as a combined Validation Set.
To purchase standards in North America or Canada, contact GE Analytical Instruments by phone at
800.255.6964 or 303.444.2009. To purchase standards in the United Kingdom, contact GE Analytical Instruments
at 44(0) 161866 9337. instruments.com. To purchase standards in Europe, Asia, and other parts of the world,
contact your local GE Sales Partner listed at www.GEInstruments.com
Standards used in each of the system protocols are shown in Table 8 through Table 11. Standards
concentrations are shown in ppm and ppb. Note that 1 ppm = 1 mg C/L, and 1 ppb = 1 µg C/L. For configurations
without conductivity measurement, standards sets without conductivity standards are available.
GE Analytical Instruments ©2010
Table 8: Standards Required for Single-Point Calibration
Calibration Blank (2 vials)
1.5 ppm TOC (as KHP)
25 µS/cm Conductivity (as HCl)
Table 9: Standards Required for Multi-Point Calibration
Calibration Blank (3 vials)
1.5 ppm TOC (as KHP)
25 µS/cm Conductivity (as HCl)
1.0 ppm TOC (as KHP)
500 ppb TOC (as KHP)
.
114 of 226
DLM 74001-04 Rev. A