Chapter 3: Installation
6.
Remove the rubber inlet cover from the top of the DI water cartridge.
7.
Slide the water bottle nozzle into the inlet hole. Make sure the tubing that connects to Port "B" is
positioned higher than the DI water cartridge, and then gently squeeze the water bottle. Be sure to fill
the reservoir until the water reaches the Full line.
8.
Connect the fitting labeled "D" to the barb labeled "B" on top of the DI water cartridge.
9.
Slide the DI water cartridge into the bracket and rotate it into the retaining clip until it snaps into place.
The cartridge must be level as you rotate it inward, and the bottom of the cartridge must be
perpendicular to the bracket.
10. Visually check for leaks from the reservoir after the DI water cartridge has been filled.
11. Turn on the Analyzer using the main power switch. This will start the DI pump. Check for large air
bubbles in all the tubing that connects to the DI water cartridge. If large air bubbles are present, tap or
manipulate the tubing to work the air bubbles out.
12. Keep the Analyzer powered on for 2 minutes.
13. After 2 minutes, turn the Analyzer off.
14. Lift the spring clip at the top of the DI water cartridge, and rotate the cartridge clockwise (as viewed
from the top) so it slides out of the retaining hooks.
15. Remove the John Guest fitting labeled "B" from the barb on the top of the DI water cartridge.
16. Connect the fitting labeled "B" to the barb labeled "B" on the top of the cartridge.
17. Connect the fitting labeled "D" to the barb labeled "D" on the top of the cartridge.
GE Analytical Instruments ©2010
Figure 9: Removing the John Guest Fitting