Chapter 4: Basic Analyzer Operation
Limits. Limits will trigger if any of the TOC or conductivity tests fail. TOC Limit only if any of the TOC tests fail, and
Cond Limit only if any of the conductivity tests fail.

Selecting the Program Language

If you have loaded a second program language into the Analyzer, you can switch between the two languages by
following these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Advanced Setup button.
4.
Press the Select Language button. Two buttons will display; the choices depend on which languages
you have loaded into the Analyzer. Select the desired language to continue.
Note that if only one program language is loaded, the Select Language button does not display.
Note
If you change the program language by mistake and need help navigating the menus to
switch back to your native language, see "Problems Changing the Program Language" on
page 165.

Configuring the Flow Sensor

In configurations with an iOS or Super iOS, the Analyzer automatically detects the presence of the sample
stream when it is working in On-Line mode. When the water flow is interrupted, the Analyzer issues an error and
stops analysis. To change the Flow Sensor setting, follow these steps:
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Advanced Setup button.
4.
Press the Flow Sensor button.
5.
By default, the Flow Sensor is set to On. To change the Flow Sensor Status, press the Flow Sensor
button and then select On or Off.
Note:
Every time the Analyzer is stopped and re-started in On-Line mode, the Flow Sensor
automatically resets to On, to protect the Analyzer.
Press the Flow Restart button and set it to On or Off. Setting the Flow Restart to On means the Analyzer
will automatically begin analysis again after a temporary stoppage of sample flow, if it was taking
GE Analytical Instruments ©2010
93 of 226
DLM 74001-04 Rev. A