High Blank Toc Warnings/Errors - GE Sievers 500 RL Operation And Maintenance Manual

On-line total organic carbon analyzer
Hide thumbs Also See for Sievers 500 RL:
Table of Contents

Advertisement

Chapter 6: Calibration and System Protocols

High Blank TOC Warnings/Errors

All system protocols that contain a Blank vial will report a warning at the end of the Blank vial measurement if
the average TOC of the blank vial is higher than 100 ppb. If this warning occurs, the Analyzer will cancel an auto
restart. The Analyzer will report an error for all system protocols at the end of of the Blank vial measurement if
the average TOC of the blank vial is higher than 150 ppb, except for the JP Protocol where the limit is 250 ppb. If
this error occurs, the Analyzer will abort the analysis and cancel an auto restart.
To disable the reporting of high Blank TOC warnings and errors during system protocols, on the Advanced Cal
menu, press the Blank Limit Test button to display the Blank Limit Test screen .
GE Analytical Instruments ©2010
133 of 226
DLM 74001-04 Rev. A