Reactivating Inactivated User Accounts; Archiving User Accounts - GE Sievers 500 RL Operation And Maintenance Manual

On-line total organic carbon analyzer
Hide thumbs Also See for Sievers 500 RL:
Table of Contents

Advertisement

Chapter 5: Password Protection and DataGuard
4.
Select the appropriate User ID from the list using the scroll buttons and press OK.
5.
Press the Password button.
6.
Specify a new Password and press Enter.
7.
Verify the new Password and press Enter.
8.
Press the Menu button to return to the DataGuard screen.

Reactivating Inactivated User Accounts

After five unsuccessful login attempts, a User ID is suspended and set to Inactive. An administrator must
reactivate the User ID before the user can log in to the system. To reactivate a User ID, follow these steps:
1.
Log in to the Analyzer with the administrator-level User ID and Password.
2.
Select the DataGuard tab.
3.
Press the Edit User button.
4.
Select the User ID from the list using the scroll buttons and then press OK.
5.
Press the User Status button and select Active.
6.
If the user has forgotten the Password, you can assign a new Password now. Press the Password
button, specify and confirm the new Password, and press Enter.
7.
Press the Menu button to save and return to the DataGuard screen.

Archiving User Accounts

If you want to remove User IDs from the user list, you must first set the unwanted User IDs to Inactive and then
archive your unused user settings as follows:
1.
Log in to the Analyzer with an administrator-level User ID and Password.
2.
Select the DataGuard tab.
3.
Press the Edit User button.
4.
Select a User ID from the list using the scroll buttons and then press OK.
5.
Press the User Status button and select Inactive.
6.
Press the Menu button to save the change and return to the DataGuard screen.
7.
Repeat steps 3 through 6 for each User ID you want to remove.
8.
Press the Archive Users button. You will be warned that user accounts set to Inactive will be removed.
Make sure you have a USB storage device attached to the Analyzer.
9.
Press the OK button.