Latching Alarms; Fhr High Alarm - GE Corometrics 170 Series Operator's Manual

Fetal monitor
Hide thumbs Also See for Corometrics 170 Series:
Table of Contents

Advertisement

Latching Alarms

FHR High Alarm

5-10
Fetal Heart Rate Monitoring: Fetal Heart Rate Alarms
Fetal heart rate threshold alarms are "latching." This means that a clinician must
acknowledge the alarm using the monitor's
alarm.
Active Threshold Alarm: Press the
component of an active threshold alarm. The visual indications remains present
until the FHR value returns to within the defined acceptable range.
Unsilenced, Resolved Threshold Alarm: If a threshold alarm condition
resolves, prior to being silenced (clinical acknowledgment), the visual and
audible indications both remain present. Press the
cancel both the audible and visual indications.
The simplest example of a high FHR alarm occurs when the FHR value is
continuously greater than the threshold (high limit) for 5 minutes. When data
consistently violates the limit, the time-to-alarm is 5 minutes. See
FHR (BPM)
1
170 Series Monitor
2003023-001
Alarm Silence
Alarm Silence
alarm activates
3
4
5
2
Figure 5-5. High FHR Alarm Example
button in order to clear the
button
to cancel the audio
button
to
Alarm Silence
Figure
5-5.
high alarm limit
(threshold)
Time (minutes)
Revision D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents