Data Transfer - Honeywell BW MicroDock II Quick Reference Manual

Automatic test and calibration station
Table of Contents

Advertisement

MicroDock II
Quick Reference Guide
To calibrate a detector, complete the following:
1.
From the docking module of the applicable detector,
press C CALIBRATION.
2.
The gas is automatically applied.
3.
When the calibration is complete, the system purges.
The countdown for time remaining displays beside
Purge.
= calibration has passed
= calibration has failed
If any test fails, refer to Troubleshooting in the Micro-
Dock II Base Station User Manual.
14
4.
The detector deactivates.

Data Transfer

(GasAlert Extreme, GasAlertMicroClip / GasAlertMicroClip XT,
GasAlertMax XT / GasAlertMax XT II, GasAlertQuattro only)
This feature transfers datalogs from the detector to the base
station MultiMediaCard (MMC).
To transfer a datalog, complete the following:
1.
Activate the detector and insert it into the docking
module.
2.
Press C DATA TRANSFER. The RUN LED flashes
rapidly. The base station LCD displays the following
screen:
Depending upon how many logs are stored on the
MMC, the data transfer may require several minutes
to complete (1-2 minutes for the GasAlertMicroClip /
GasAlertMicroClip XT, GasAlertMax XT / GasAlert-
Max XT II).
The base station MMC temporarily stores the datalogs. To
save the datalogs for future use, transfer the files to Safety
Suite Device Configurator (SSDC). Refer to the SSDC
Operator's Manual.
Note