Honeywell BW MicroDock II Quick Reference Manual page 15

Automatic test and calibration station
Table of Contents

Advertisement

The LCD displays the results of the bump test. The
PASS LED lights green.
If any test fails, the FAIL LED lights red. Refer to
Troubleshooting in the MicroDock II Base Station
User Manual.
3.
Press C OK to return to normal operation.
4.
After 5 minutes of inactivity, the detector deactivates.
For additional information, refer to the MicroDock II
Base Station User Manual.
Calibration (excluding GasAlertClip Extreme)
Calibration
(excluding GasAlertClip Extreme)
a
BW recommends using premium grade calibration gases
and cylinders that are certified to National Standards. The
calibration gases must meet the accuracy of the detector.
All calibration cylinders must be used with demand flow
regulators and must meet the following maximum inlet
pressure specifications
• Disposable cylinders 0-1000 psig/70 bar
• Refillable cylinders 0-3000 psig/207 bar
a
Ensure the detector is not in a low battery state prior to
performing a calibration.
A calibration adjusts the sensor sensitivity to ensure an accu-
rate response to gas. During the calibration, event logs from the
GasAlertClip Extreme, GasAlert Extreme, GasAlertMicroClip /
GasAlertMicroClip XT, GasAlertMax XT / GasAlertMax XT II,
and GasAlertQuattro are transferred to the MMC.
The base station cannot calibrate Cl
sensors.
MicroDock II
Warning
Caution
Note
, O
, and ClO
2
3
2
13