User Options Menu - Honeywell BW MicroDock II Quick Reference Manual

Automatic test and calibration station
Table of Contents

Advertisement

MicroDock II
Quick Reference Guide
Pushbuttons
Pushbutton
C
BUMP CHECK
C
CALIBRATION
C
DATA TRANSFER
4
Description
To bump test a detector, press
BUMP CHECK
.
When connecting a new docking
module, press and hold
C BUMP CHECK
to send a
confirmation signal back to the
base station.
To calibrate a detector, press
C CALIBRATION
(all models
excluding GasAlertClip Extreme).
To transfer data from a detector,
C DATA TRANSFER
press
(GasAlert Extreme, GasAlertMi-
croClip / GasAlertMicroClip XT,
GasAlertMax XT / GasAlertMax
XT II, and GasAlertQuattro only).
The Automatic Datalog Down-
load option is available for the
GasAlertMicroClip / XT, GasAlert-
Max XT / II, and GasAlertQuattro
docking modules only. For more
Data Trans-
information, refer to
fer.

User Options Menu

Description
Time/Date modifies the time and
C
date of the base station.
Inlet Setup modifies the gas
type, the gas concentration lev-
els, and the gas cylinder lot num-
bers.
Pump Setup modifies the station
pump speed. Recommended
pump speeds are 60% (350 ml/
.
min.). The pump speed must be
set for each new docking mod-ule
that is added to the station.
NOTE: A flow meter is required.
Contrast brightens or dims the
text of the LCD.
Backlight
enables/disables the
LCD backlight. When enabled,
auto
the
option automatically
deactivates the backlight when
the base station is not in use.
Display