Honeywell BW MicroDock II Quick Reference Manual page 10

Automatic test and calibration station
Table of Contents

Advertisement

MicroDock II
Quick Reference Guide
station LCD displays the docking module number, and
the type and serial number of the detector.
GasAlertMicro
• Insert the bottom of the detector (serial number face up)
into the detector bay first and then lower the top into
place.
• Close the lid and press until the release tabs click.
• When the detector has been inserted correctly, the RUN
LED(s) on the docking module light yellow and Unit
Inserted displays on the base station LCD. The base
station LCD displays the docking module number, and
the type and serial number of the detector.
GasAlertMicro 5/PID/IR
Important: If the GasAlertMicro 5/PID/IR detector is fitted with
a pump module, the diffusion adapter must be removed from
the docking module. Refer to the MicroDock II Base Station
User Manual.
• Insert the GasAlertMicro 5/PID/IR (LCD facing up) at a
45° angle and insert the bottom into the docking bay.
Ensure the connector outlets on the bottom of the detec-
tor lock into place over the connector pins in the docking
bay.
• Close the lid and press until the release tabs click.
8
• When the detector has been inserted correctly, the RUN
LEDs light yellow, Unit Inserted displays on the base
station LCD, and MicroDock displays on the detector
LCD. The base station LCD displays the docking mod-
ule number, and the type and serial number of the
detector.
GasAlertMicroClip / GasAlertMicroClip XT
Important: If the GasAlertMicroClip / GasAlertMicroClip XT is
fitted with a calibration cap or an auxiliary filter, it must be
removed prior to inserting it into the docking module. Refer to
the GasAlertMicroClip User Manual or GasAlertMicroClip XT
Technical Reference Guide and the MicroDock II Base Station
User Manual.
The GasAlertMicroClip cannot be used in the
GasAlertMicroClip XT docking module and vice
versa.
• Insert the bottom of the detector (serial number face up)
at a 30° angle into the detector bay.
The docking module lid only raises upward 30°. Do
not force the lid beyond its limit.
• Close the lid and press until the release tabs click.
• When the detector has been inserted correctly, the RUN
LEDs on the docking module light yellow and Unit
Inserted displays on the base station LCD. The base
Note
Note

Advertisement

Table of Contents
loading

Table of Contents