Message
b) Warnings:
Data restore
Outp.1 Error
Outp.2 Error
I1-Warning
I2-Warning
Update error
Softw. error
Sett. error
Description, process
DL230 batteries failed intermittently. As a result of this,
the time has not changed and measurement acquisition
has failed. All data has been saved.
This message appears when the battery is changed and
the old batteries are removed before the new ones are
connected. (see chapter 7.2, page 76).
Process using parameterization program "enSuite":
Set the DL230 clock
Clear the status register
More pulses need to be sent to the named output, than
its programming permits.
To eliminate the cause of this problem, use the "enSuite"
parameterization program to
either reduce the cp value of the output
or increase the output frequency.
In both cases, the change must be aligned with the de-
vice connected to the DL230.
The encoder could not be read at the end of the meas-
urement period. The flow recording at this point is invalid.
An error in the cached software image was found prior to
the software update.
This message is used for factory diagnostics.
Please contact the Elster customer service team (see
chapter 1.3 "Customer service").
Due to the programming carried out an unusable settings
combination has been created.
Please contact the Elster customer service team
(see chapter 1.3 "Customer service").