Assembly, Connection and Putting into Operation
Use only cables, which fit to the cable glands in order to
keep IP protection class also after the installation.
Minimum cable diameter:
Cable gland M12 ≥ 5 mm
Cable gland M16 ≥ 5,5 mm.
Different downstream devices can be connected to the
digital outputs of the DL230. The outputs are preconfigured
for this purpose (see chapter 5.3.1.5).
Ill.: 10
5.2.4.1 Electrical insulation of the outputs
In standard cases, all negative poles of the outputs are electrically con-
nected to the motherboard.
For special applications, i.e. switching a positive pole, each output can be
electrically separated from the motherboard and from the other outputs.
CAUTION!
Reduced battery life
Activation of electrical isolation of outputs reduces battery life
when operating the battery.
It is then impossible to give a reliable prediction of the re-
maining battery life.
1. Remove the terminal cover for the digital
outputs (terminals DA1 and / or DA2) (Ill.: 10).
2. Connect the downstream device to the
corresponding digital outputs (terminals DA1
to DA2) of the DL230 (Ill.: 10).
3. Screw the terminal covers over the connected
terminals after connecting the cables. If
required, a seal should then be bonded to the
fastening screw (see chapter 5.3.2).
4. If necessary, adjust the cp value (pulse
contact) for the pulse outputs as described in
chapter 5.3.1.5.