Assembly, Connection and Putting into Operation

5.3 Putting into operation

5.3.1 Configuration of measurement parameters

If the DL230 is subject to calibration regulations, the works
described below should only be performed by legally
authorized individuals.
The necessary operating parameters can be adjusted using the free
configuration program "enSuite", which is available under www.elster-
instromet.com.
If the program is not available, the configuration can also be carried out
using the keyboard as described below (see chapter 5.3.1.2).
Before adjusting the measurement parameters via the
keyboard, please read chapter 6 to learn how the device is
operated.
5.3.1.1 Opening the calibration lock
The calibration lock is located at the back of the housing cover in the form
of a button, and this can be secured by means of an adhesive label. This
button must be pressed in order to adjust the values and parameters pro-
tected by configuration regulations.
Ill.: 12
43
Position of the
push-button to
open the calibra-
tion lock.