Installation And Service Checkout Procedures; Mechanical Checks - GE PANDA Service Manual

Warmers
Table of Contents

Advertisement

Chapter 2: Installation, Maintenance, and Checkout

Installation and Service Checkout Procedures

2.1.2.3 Service tools
Heater dish alignment laser kit (includes holder) (takes
three #392 1.55V button cells or equivalent)
Laser holder (includes O-ring)
NULL Modem Cable
Light gray touch-up paint (Munsell .16GY8.56-0.44
chroma), 18 ml
Masimo SET 1 electrical leakage test cable
Masimo SET 2 electrical leakage test cable
Nellcor electrical leakage test cable
Crank handle kit
Note:
The resuscitation systems service tools are described in the Resuscitation Systems service manual.
Note:
A straight-line laser level (with two wall leveling capability) or a combination of T-square and Plumb-
bob is recommended for accurate heater alignment of Freestanding and Wall Mount Warmer
models.
2.2 Installation and Service Checkout Procedures
Warning: Use only the specific power cord supplied with the unit. Use of an
unapproved power cord could result in damage to the unit or injury.
Except for Panda Wall Mount Warmer model, all other warmer models do not require any installation. After
removing the warmer unit from the shipping package, perform the following checkout procedures before
putting the unit into use.
For Panda Wall Mount Warmer model, refer to the installation instructions in
secure the unit on the wall and do heater alignment. Then perform the following checkout procedures
before putting the unit into use.

2.2.1 Mechanical Checks

Warning: Do not perform the preoperational checkout procedure while the patient
occupies the unit.
Warning: Complete the preoperational checkout procedure section of this manual
before putting the unit into operation. If the equipment fails any portion of the
checkout procedure it must be removed from use and repaired.
1.
Disconnect the power cord for the mechanical portion of the preoperational checkout procedure.
2.
Examine the power cord for any signs of damage. Replace the cord if damage is evident.
3.
Check that the cord retainer/accessory outlet guard on the back of the control panel is in place.
4.
Examine the unit overall for any damaged or missing parts.
5.
Examine the unit for any missing or damaged labels. Make sure the labels are attached in the proper
locations. For a list of labels and their proper locations, refer to
2-2
M1118361
M1118362
M1128795
6600-0714-200
2006036-001
M1230051
M1237129
M1163421
Appendix E
section
6.11.
to mount and
Service Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Giraffe