GE PANDA Service Manual page 186

Warmers
Table of Contents

Advertisement

Appendix C: Alarm Priorities
Activation
Priority Alarm Subject
Criteria
7
Baby
>2.0° from
temperature
Control Temp
8
Check baby
Radiant %
power @ 100%
for > 12 min
9
Check baby
Radiant Heater
% above
"Check Baby
Alarm
Disabled"
(CPAD) limit for
> 12 min
10
Temperature
Two
probe failure
Thermistors in
a probed differ
by 0.5° C or
more.
11
Disconnected
No longer
temperature
getting a
probe
reading from
the
temperature
probe
12
Baby
>1.0°C (>0.5°C)
temperature
from control
temp
13
Baby
<1.0°C (<0.5°C)
temperature
from control
temp
C-2
Sound:
1 = High
Message
Silence
2 = Low
Displayed
Time
3 = informational
Baby Cold-
5 min
Check
Temp
Probe
Check
12 min
Baby
Check
12 min
Baby
Temp.
2 min
Probe
(after 30 secs.)
Failure
Confirm
2 min
Probe Jack
(after 30 secs.)
Connection
Baby Hot -
12 min
Check
Temp
Probe
Baby Cold -
12 min
Check
Temp
Probe
Notes/
Comments
1
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
2
2
2 to 1
2 to 1
2
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
2
With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Indication
Special
Light
Requirements
Primary
Operational in
Baby Control
Mode
Primary
Operational in
Baby Control
Mode
Primary
Operational in
Baby Control
Mode
Primary
Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
Primary
Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
Primary
Operational in
Baby Control
Mode
Primary
Operational in
Baby Control
Mode
Service Manual