Mattress Height Label Installation (For Installations With Bassinets - GE PANDA Service Manual

Warmers
Table of Contents

Advertisement

Appendix E: Installing Wall Mount Units

Mattress Height Label Installation (for Installations with Bassinets)

E.5 Mattress Height Label Installation (for Installations with Bassinets)
1.
Install one mattress height label on the wall as shown in
(indicating the minimum height) at 32" +/- .1" (81.30 cm +/- .25 cm) from the floor and vertically center
align the label with the center of the warmer head.
2.
Install one mattress height warning label next to the mattress height label on the wall.
E-10
MATTRESS
HEIGHT
WARNING
LABEL
32.0" ± 0.1"
81.30 cm ± 0.25 cm)
FIGURE E-7. Mattress Height Label Installation
Figure
E-7. Place the bottom bar of the label
MATTRESS
HEIGHT
LABEL
BUMPER