User Manual
8.4
General
Fault Symptom
Unit will not
connect / write
to USB Memory
Stick
Unit will not
connect to
calibration
management
software
Unit will not
Switch On
Unresponsive
Unit
Page 104 of 105
Possible Cause
USB flash drives with a
capacity of 1GB can take a
few seconds to fully connect
Communications mode set
to storage
Faulty / Flat battery
Processor Lock-up
Check
Try with smaller capacity USB Drive
From the DASHBOARD select
SETTINGS then DEVICES; the USB
client port should be set to
Communications.
Try Using External Power Supply
Remove Battery and Replace
Press and hold power switch
Remove Battery / Power and replace
[EN] English – 109M4017 Revision -
Druck DPI612