GE Druck DPI612 User Manual page 45

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
Table 2-4
Step
Procedure (Vacuum)
1
Set to vacuum operation
(-).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
4
Use the pump to set the
maximum vacuum or set
the vacuum you want to
adjust.
5
Adjust the vacuum
(+ decrease; - increase).
Page 45 of 105
Table 2-5
Step
Procedure (Pressure)
1
Set to Pressure operation
(+).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster fully
counter clockwise.
3
Seal the system.
4
Use the pump to
pressurise the system (up
to ≈20 bar)
5
Open the refill valve (1
turn)
6
Increase the pressure
with the volume adjuster
7
If increasing pressure the
limit of travel is reached,
close the refill valve
8
Wind the volume adjuster
fully counter clockwise.
There is no change in
pressure.
9
Refill the pressure
mechanism with the
pump (≈15 cycles)
10
Wind the volume adjuster
clockwise until the
pressure starts to
increase
11
Continue to do steps 7 to
10 until you get the
necessary pressure.
[EN] English – 109M4017 Revision -