DOCUMENTATION

This chapter describes the Documenting functions available with the
Druck DPI612 calibrator as follows:
4.1

ANALYSIS

The Analysis function takes readings from two or more DPI612
channels to calibrate the transfer characteristic of the device being
tested. One channel is the Reference channel and is used as follows:
The other channel is the Input channel and is used as follows:
Any active channel that is not defined as Reference is Input by
default.
There must be one Reference channel and at least one Input
channel defined for the Analysis function set-up to be correct.
At each value the Analysis function calculates the difference of each
Input channel to the ideal transfer characteristic and compares this
to a tolerance limit.
Page 81 of 105
ANALYSIS
RUN PROCEDURE
It provides a measure of the input signal to the device.
If the device is a pressure transmitter the Reference
channel would be Pressure, measuring the input pressure
to the device.
Measures the output signal from the device
If calibrating a process transmitter it could be the electrical
channel in Current Measure mode.
The deviation is shown in %Span or %Rdg
The tolerance test result is shown a Pass
icon.
[EN] English – 109M4017 Revision -
Druck DPI612
or
Fail