Advanced Task

General

Advanced task allows the user to configure the instrument to monitor one of a number of
different input measurements and outputs (sources). Additionally, five process functions,
Tare, Max/Min, Filter, Flow and % Span can be applied to the input functions.

Select Input

To select an input channel, select ADVANCED task from the task menu. The display
shows the list of the input selections and, if available, the PROCESS soft box (F1) and the
UNITS soft box (F2).
The following procedure shows the method of input channel selection and the method of
changing units:
Note: Left/right arrow keys function as page up/down keys.
Refer to pages 20 to 23 for details of process functions.

Ambient Temperature Measurement

To set-up the instrument to read ambient temperature, proceed as follows:
INPUT
SELECT INPUT
Note: Ensure that the temperature reading has stabilised.
SELECT INPUT
AMBIENT TEMP
NO INPUT
PROCESS
UNITS
SELECT AIR
TEMPERATURE
UNITS
CELSIUS
FAHRENHEIT
19
Select Input
SELECT UNITS
K0430 Issue No. 2