Procedure
Step
1
2
3
4
5
Page 53 of 105
1
If applicable, release the pressure (Section 2.3.6) and remove
the device (Section 2.1.1).
Remove the hydraulic pressure release valve.
Wind the volume adjuster wheel fully clockwise; this moves
the fluid out of the pressure mechanism
Put a container below the instrument then tilt the instrument
up until all the fluid has come out. To discard the hydraulic
fluid, obey all the local health and safety procedures.
Note: Fluid comes out of the test port and the connection for the
pressure release valve.
To flush out fluids that contain unwanted material, refill the
system and repeat steps 3 and 4.
Note: To prevent contamination use only one type of hydraulic fluid
in the instrument.
2
3
Procedure
[EN] English – 109M4017 Revision -
Druck DPI612
4