Vacuum Or Pressure Operation Dpi 612 Pflex (20 Bar); Release Pressure Dpi 612 Pflex (20 Bar) - GE Druck DPI612 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI612:
Table of Contents

Advertisement

User Manual
2.2.3

Vacuum or Pressure Operation DPI 612 pFlex (20 bar)

1
Table 2-2
Step Procedure (Vacuum)
1
Set to vacuum operation
(-).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
4
Use the pump to set the
maximum vacuum or set
the vacuum you want to
adjust.
5
Adjust the vacuum
(+ decrease; - increase).
2.2.1

Release Pressure DPI 612 pFlex (20 bar)

Page 43 of 105
2
Figure 2-5 Vacuum or Pressure operation
Open the pressure release valve counter clockwise (1 turn)
3
4
Table 2-3
Step
Procedure (Pressure)
1
Set to Pressure operation
(+).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
4
Use the pump to set the
maximum pressure or set
the pressure you want to
adjust.
5
Adjust the pressure
(+ decrease; - increase).
[EN] English – 109M4017 Revision -
Druck DPI612
5