PROCEDURES: DC Volts (source)

Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1).
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a two-point
calibration for the applicable range:
Page 91 of 105
Table 5-4 Millivolts (measure) error limits
mV
±2000
±1000
±200
±100
Table 5-5 Voltage (measure) error limits
V
±30
±21
±20
±10
±5
0
Volts (source) range
Calibrator
uncertainty
(mV)
0.051
0.040
0.051
0.0040
0.0036
Calibrator
uncertainty
(V)
0.00052
0.00040
0.00031
0.00016
0.00008
0.000024
10 V
24 V
[EN] English – 109M4017 Revision -
Druck DPI612
Permitted
DPI612 error
(mV)
0.1280
0.0940
0.0148
0.0110
0.0070
Permitted
DPI612 error
(V)
0.00180
0.00154
0.00118
0.00089
0.00075
0.00060