Part 7: Approvals Informatio; Electromagnetic Compatibility (Emc); Fcc - Omega iServer Microserver User Manual

Table of Contents

Advertisement

PART 7
APPROVALS INFORMATION
7.1 CE APPROVAL
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC,
and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Basic Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min
• Input Power to Ethernet Output:
• Input Power to I/O lines:
• Input Power to I/O lines:
• Ethernet Output to I/O lines:
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains
Supply (power).
Transients Overvoltage Surge (1.2/50uS Pulse)
• Input Power:
• Ethernet:
Note:
The ac/dc power adaptor must have Safety Qualified Agency Approvals
for CE with Double Insulation rating.
The ac/dc power adaptor is 9Vdc.
The minimum output current rating is 500mA.
EMC EN61000-6-1:2001 (Immunity) and EN61000-6-3:2001 (Emmissions)
Immunity requirements for residential, commercial and light-industrial environments
• EMC Emissions
• EMC Immunity
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control
and laboratory.
• EMC Emissions
• EMC Immunity
Note:
I/O lines require shielded cables and these cables must be located
on conductive cable trays or in conduits.
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for
additional information.
7.2

FCC

This device complies with Part 15, Subpart B, Class B of the FCC rules.
1500Vac (iFPX-W and iFPX-D)
none (iFPX-W)
1500Vac (iFPX-D)
1500Vac (iFPX-D)
500V Transients Overvoltage
1500V Transients Overvoltage
Table 1, Class B
Table 1: Enclosure
Table 2: Signal Lines Ports
Table 3: Dc input/Dc output Ports
Table 4, Class B of EN61326
Table 1 of EN61326
66