Figure 19: Velocity Method - Installation; Figure 20 Velocity Method - Adjusting The Sipper Tube Height - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 4: Installation
5.
Close the two clamps over the pipe and use pliers or channel locks to secure.
6.
Adjust the height of the sipper tube to enable drawing sample from a laminar flow without bubbles.
7.
Connect a section of inlet tubing (supplied in Kit) from the bulkhead to the fitting on the top of the sipper
tube.
8.
Test the proper adjustment of the sipper tube by performing a sample flush and verifying that no
bubbles are pulled through the clear inlet tubing.
9.
Repeat steps
3
10. If you are also installing the fail safe option, cut and remove the tie wrap used to secure the sensor to
the Analyzer during shipping.
11. Remove the U-bolt from the sensor and position the U-bolt over and near the end of the pipe with the
sipper tube.
GE Analytical Instruments ©2016
Figure 19: Velocity Method — Installation
Figure 20: Velocity Method — Adjusting the Sipper Tube Height
through
8
for each stream.
To Waste Outlet
Pipe Nearest to Instrument (with internal sipper tube)
To Sample Inlet Source
Pipe Farthest from Instrument
93 of 270
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents