Gas Leak Checks - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 9: Troubleshooting
To perform the Liquid Reactor Flow test
Press the Liquid Reactor Flow button. The button appears orange and the Analyzer displays the various Test
states.
Draining — Pump P1-R empties any liquid that may be in the sparger through the reactor bypass in
preparation for the test.
Loading —The syringe fills with dilution water, which is then dispensed into the sparger. Valve V6-R is
closed and valves V3-R, V4-R, and V5-R are open.
Dispensing — Pump P1-R empties the sparger through the reactor tube.
Reactor Flow — It will take a few seconds for the water to exit the Liquid Waste tube located in the back
of the Analyzer.
If you do not see water exiting the waste tube while P1-R is operating, or if the liquid level in the sparger
does not go down, the reactor tube may be plugged.
Waiting — The test is complete.

Gas Leak Checks

There are six tests to help locate leaks in the gas delivery system that may exist in the Analyzer. Step-by-step
instructions for each leak test is provided in the following pages. Perform the procedures in the following
sequential order.
Check 1 — Gas Manifold Leak Check (using the Gas Manifold button)
Check 2 — Reactor Inlet Valve Leak Check (using the Reactor Inlet Valve button)
Check 3 — Reactor Bypass Leak Check (using the Reactor Bypass button)
Check 4 — Reactor Leak Check (using the Reactor button)
Check 5 — Reactor Outlet Valve Leak Check (using the Reactor Outlet Valve button)
Check 6 — Gas-Liquid Separator, NDIR Filter, NDIR Sample Path and Sparger Leak Check (using the
Reactor Bypass button)
Before You Begin
Before starting the Gas Leak Checks, ensure that either an optional air pump is installed or the gas inlet to the
Analyzer is connected to a source of compressed gas (10-20 psig). Refer to
information.
Also, make sure to insert the small plug (included in the Accessories Kit) into the end of the coiled tube coming
from the left side of the NDIR. This prevents leakage of vent gas out of the NDIR during testing.
GE Analytical Instruments ©2016
Chapter 3, "Interferences"
226 of 270
for more
DLM 68100-09 EN Rev. A

Advertisement

Table of Contents
loading

Table of Contents