Chapter 8: Maintenance
7.
Remove the syringe from the Analyzer by pulling it down and away from the valve.
The washer should remain in the valve. If the washer comes free of the valve when you remove the
syringe, be sure to capture the washer for re installation. You must place the washer back on the
plunger tip when reinstalling.
8.
Pull the plunger all the way out of the syringe barrel.
GE Analytical Instruments ©2016
Figure 77: Removing the Syringe
Figure 78: Pulling Out the Plunger