Chapter 2: System Description

Additional System Specifications

Carrier Gas
Sample Analysis Volume

Operating Specifications

Sample Time (min)
(Time to reported value)
Sample Consumption
(See aliquot in respective range)
Monthly Carrier Gas
Electrical
(analysis complete)
GE Analytical Instruments ©2016
CO
free air or Nitrogen; Gas must be oil- and water-free;
2
Delivery Pressure of 50-100 psi required when using the air purifier or
Carrier Gas CO
Scrubber options; Delivery pressure must be regulated
2
to 30-100 psi when using a cylinder of purified gas; Unit will consume
up to 200 cc/min in NPOC mode.
Range
Up to 100 ppm
Up to 1,000 ppm
Up to 5,000 ppm
Up to 20,000 ppm
Up to 50,000 ppm
Rep 1
Rep 2+
Sample 1 Flush
Sample 2 Flush
Dilution Flush
Note: The time for the analytical cycle is the same whether or not the
Sample Backflush feature is utilized.
Ox OL Sample
Filter Panel
Temperature
Fast Flush
Sample Flush
Supply Pressure
Consumption
Note: The consumption rate is not relevant to the compressed air needed
to operate the Regenerative Air Purifier option. In addition, this flow rate
is not sufficient to operate the IP-56 option. GEAI does not recommend
using cylinder air for the IP-56 option.
Power
Alarm Relays
Binary Input
Binary Output
Analog Output
Ethernet
53 of 270
Sample Volume
2.82 mL
2.00 mL
0.80 mL
0.20 mL
0.08 mL
NPOC (OL)
8.3
5.8
+ 1.4 per rep
+ 5.2 per rep
+ 5.2 per rep
+ 0.7 per rep
50 mL/min and  125 psig
>1.0 L/min w/o exterior sample pump
400 mL/min with exterior sample pump
60 °C
10 mL
Refer to aliquot volumes in respective range.
30 to 100 psig
200 cc/min in NPOC mode, Less in TOC mode.
Standard cylinder holds 300 cu ft, or 30-days
worth
100 - 240 VAC +/-10% 50/60 Hz, 400 VA
30 VDC @ 1A max or 125 VAC @ 0.5A max
5-24 V (remote start)
Relay rated max 125 VAC or 110 VDC @ 1A
4-20 mA, max load @ 600 ohms, powered
Modbus port 502
TOC & TC (OL)
9.7
8.3
+ 1.4 per rep
+ 0.7 per rep
DLM 68100-09 EN Rev. A