Chapter 2: System Description
Operating Specifications
Analysis Modes
Range
Precision
Accuracy
Linearity
LOD
TOC Calibration Stability
Analysis Time
Particle size
Sample Temperature
Ambient Temperature
Source Sample Pressure
Minimum Sample Flow Rate
Drain
Analyzer Specifications
Outputs
Inputs
Display
Power
Alarm Relays
Communications Protocols
Installation/Overvoltage
Category
Sample Streams
Dimensions
Weight
Safety Certifications
Environment
Enclosure Options
Maximum Relative Humidity
Maximum Altitude
Pollution Degree
________________________________
1
Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
GE Analytical Instruments ©2016
1
RSD ≤3% of reading at >5 ppm NPOC or TOC
Greater of ±
3% of reading or
R
2.6 to 8.3 minutes in Online mode
≤200
μm without filtration or >200 μm with optional wastewater sampler
10-60q C (41-140 qF),
or 10-85 qC (41-185 qF) with high-temperature options
,
125 psig maximum with provided control valve
0.5 psig minimum without optional sample pump
50 mL/min without optional sample pump
External USB Port (1); Internal USB Ports (2); Ethernet (1);
Binary End-of-Analysis Output (1);
Isolated 4-20 mA Analog Outputs (5 analytical; 1 for stream ID)
Binary Input (1) for Remote Analysis Start/Stop
100 – 240
System Fault Alarm (1); User-Configurable Alarms, High and Low Limit (4)
2 Streams (standard), 5 Streams (optional)
H: 92.7 cm (36.5 in.); W: 64.6 cm (25.4 in.); D:38.7 cm (15.3 in)
CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 NO. 61010-1
IP56/NEMA 4X/Class 1 Division 2/ATEX
52 of 270
InnovOx
On-Line TOC Analyzer
NPOC, TOC (TC-IC), TC, IC
0.250 to 50,000 ppm TOC
±
0.25 ppm, 1 to 100 ppm, NPOC or TOC
2
≥ 0.995, measured as NPOC
250 ppb NPOC
1
Up to 6 months
10-40 qC (50-104 qF)
Gravity Drain
Color LCD w/touch-screen
±
10% VAC, 50 – 60 Hz, 400 VA
Modbus
®
; TCP/IP over Ethernet
II
36.3 kg (80 lb)
Up to 95%, non-condensing
3,000 m (9,800 ft)
2
DLM 68100-09 EN Rev. A