Chapter 5: Calibration and Verification

Point-to-Point Multi-Point Calibration

The advantage of selecting the point-to-point calibration is that any non-linearity of response is compensated
for during the calibration.
Figure 35
depicts a point-to-point calibration with 6 points. Like the linear fit, the point having the lowest
concentration is measured first, and the points must increase in concentration at each step. Point 1 can either
be reagent water or a higher, known concentration.
The Analyzer calculates an intercept and slope for the line connecting each pair of points, as shown in
When this calibration is later applied to a sample, the Analyzer applies the appropriate intercept and slope,
based on the magnitude of its mass response for that sample.
GE Analytical Instruments ©2016
Figure 34: Linear Fit Six-Point Calibration
112 of 270
Figure
35.
DLM 68100-09 EN Rev. A