Electrical Safety - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

Direction 5936427, Rev. 1
1-5-4

Electrical safety

Safe practices
RISK OF DAMAGE TO EQUIPMENT
CAUTION
CONNECTING THE LOGIQ TOTUS TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY IT.
RISK OF DAMAGE TO EQUIPMENT
CAUTION
USE THE POWER CORD EXCLUSIVELY ON THE LOGIQ TOTUS. DO NOT USE THE POWER
CORD TO MAKE ANY OTHER ELECTRICAL CONNECTIONS.
Follow these guidelines to minimize shock hazards whenever you are using the LOGIQ Totus:
To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
The LOGIQ Totus is equipped with a three-conductor AC power cable. This must be plugged into
an approved electrical outlet with safety ground. A separate power outlet with a 20 amp circuit
breaker for 120 VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or
15 amp circuit breaker for 100 VAC for Japan.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the LOGIQ Totus power cable and the power connector must meet international electrical
standards.
Probes
Follow these guidelines before connecting a probe to the LOGIQ Totus:
Inspect the probe prior to each use for damage or degradation to the:
-
-
-
-
-
-
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any liquid.
The LOGIQ Totus has more than one type of probe port. Use the appropriate probe port designed
for the probe you are connecting.
Peripherals
Refer to the "Peripherals used in the patient environment" in Chapter 2 of the User Manual for
peripheral isolation information.
GE HealthCare CONFIDENTIAL
housing
cable strain relief
lens
seal
connectors
locking mechanism
Chapter 1 Introduction
LOGIQ Totus Basic Service Manual
1 - 21