Switch - Honeywell BW RigRat User Manual

Local area gas monitor
Hide thumbs Also See for BW RigRat:
Table of Contents

Advertisement

Switch

BW RigRat supports three separate SPST PhotoMOS relay output switch connections for
applications where other equipment needs to be controlled (lights, sirens, etc.).
Relay Definitions:
Relay 1
Any gas sensor failed, or STEL/TWA/+OL/-OL alarms
Relay 2
Any high alarms
Relay 3
Any low alarms
IMPORTANT! This is a fixed configuration that can only be modified via Device Configurator
app.
Note: Driving high current or voltage through the switch will damage the product.
Note: The default option is NO (normally open). It can be set to NC (normally closed) via the
Device Configurator app. Note that NC consumes more battery current.
Voltage
Max power consumption on switch
Continuous load current
On resistance
BW RigRat
PhotoMOS relay output switch parameters
Must meet IS parameters if used in hazardous area!
Item
Value
30V
0.75W
100mA
1Ω typical
56
Note
AC/DC
User Manual