Replacing Sensors; Battery Replacement - Honeywell BW RigRat User Manual

Local area gas monitor
Hide thumbs Also See for BW RigRat:
Table of Contents

Advertisement

Replacing Sensors

1.
Gently lift out the desired sensor module with your fingers. If it is seated very tightly in
the socket, do not pull it with heavy force. Gently rock it slightly back and forth while
lifting.
2.
Install the replacement sensor. It can only go into its slot one way. The connector inside
the instrument and the indexing guides are good visual indicators of how to set the
sensor into position. Make sure the indexing keys are aligned and press the sensor into
place to ensure it is seated firmly.
Match index keys on sensor module with socket
3.
Replace the cover and tighten the four screws.
It is extremely important that sensors are installed in the correct orientation.
When installing a new sensor, the plastic film should be on top of the sensor. Once the sensor is
installed, remove the plastic film.
IMPORTANT! Always perform a full calibration after replacing sensors.

Battery Replacement

If the two rechargeable batteries ever need replacement, the work should only be performed by
those trained by Honeywell to service this instrument. Replacements should only be of the
same type: P/N 500-0165-000 (or 500-0165-001 for NA certification).
WARNING! Do not replace battery in hazardous locations.