Honeywell BW RigRat User Manual page 9

Local area gas monitor
Hide thumbs Also See for BW RigRat:
Table of Contents

Advertisement

RISK OF EQUIPMENT DAMAGE
Only use the Honeywell-specified rechargeable lithium-ion battery pack supplied with
l
the instrument.
Charge the instrument Li-ion battery using the specifically supplied charger and only
l
outside hazardous areas. The maximum voltage from the AC charger Um must not
exceed 6.0 VDC.
The battery pack can only be changed in a non-hazardous area. Only use approved
l
rechargeable battery pack P/N 500-0165-000 (or 500-0165-001 for NA certification)
produced by Honeywell.
Any data download device connected to this instrument must be approved SELV or
l
Class 2 equipment.
Use of non-Honeywell components will void the warranty and can compromise the safe
l
performance of this product.
Warning: Substitution of components may impair safe performance of this product.
l
When changing the sensor configuration, ensure the operator is aware of any changes
l
to certification restrictions.
CAUTION: BEFORE EACH USE, SENSITIVITY OF THE COMBUSTIBLE GAS SENSOR MUST BE
TESTED ON A KNOWN CONCENTRATION OF METHANE GAS EQUIVALENT TO 20 TO 50% OF
FULL-SCALE CONCENTRATION.  ACCURACY MUST BE WITHIN 0 AND +20% OF ACTUAL. 
ACCURACY MAY BE CORRECTED BY CALIBRATION PROCEDURE.
ATTENTION: AVANT CHAQUE UTILISATION VERIFIER LA SENSIBILITE AVEC UNE
CONCENTRATION CONNUE DE METHANE EQUIVALENTE A 20 to 50% DE LA PLEINE
ECHELLE.  LA PRECISION DOIT ETRE COMPRISE ENTRE 0 to 20% DE LA VALEUR VRAIE ET
PEUT ETRE CORRIGEE PARUNE PROCEDURE D'ETALONNAGE.
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE AN EXPLOSIVE CONCENTRATION.    
ATTENTION: DES LECTURES SUPÉRIEURES A L'ÉCHELLE PEUVENT INDIQUER DES
CONCENTRATIONS EXPLOSIVES.
BW RigRat