Table 3: SLG720 bypass and stillwell recommended diameters

Probe type
Rod probe
Wire probe
Coaxial probe

Table 4: SLG726 bypass and stillwell recommended diameters

Probe type
Rod probe
Wire probe
Coaxial probe
Chambers, bypass pipes, and still wells with smaller diameter can lead to problems
with build-up. Chambers larger than 6" (150 mm) can be used, but offer little
advantage for radar measurement.
The probe must extend the full length of the chamber and not contact the bottom of
the chamber, or make contact with the chamber wall.
Clearance from the bottom of the chamber is recommended to be 1" (25 mm). Probe
selection is dependent on length.
For lengths less than 20' 8" (6.3 m): Rod probe is recommended.
For lengths more than 20' 8" (6.3 m): Wire probe with weight and centering disk is
recommended, only wire probes are available.
A centering disk is recommended for rod probes over 1m length to prevent excessive
movement caused by strong currents inside the pipe.
22
Recommended diameter
3" or 4" (75 mm or 100 mm)
4" (100 mm)
N/A
Recommended diameter
3" or 4" (75 mm or 100 mm)
4" (100 mm)
N/A
Quick Start Installation Guide
Minimum diameter
2" (50 mm)
2" (50 mm)
1.5" (37.5 mm)
Minimum diameter
2" (50 mm)
2" (50 mm)
2" (50 mm)