Rod probe assembly

Rod probes are shipped in segments. The segments are attached with a stud and a
lock washer.
Step
Fully thread the nut onto the central conductor. Using a lock washer,
thread the first rod segment on to the central conductor.
1
Torque the nut against rod probe and lock washer to secure the connection.
Thread the stud into first rod segment end. Using a lock washer thread the
2
next segment onto the stud. Apply torque to secure the connection.
Note: Tighten each rod connection point to the following torques:
SLG720 6.0 Nm (4.4 ft-lbs)
SLG726 15 Nm (11 ft-lbs)
Note: For flanged SLG726 models, ensure the nut does not intrude into the process
connector. See
Figure 3: SLG726 flanged process connection, probe nut installation position, mm
6
Figure 3
for more information.

Figure 2: Rod probe assembly

[in] probe selection