Sony Tektronix VX4792 User Manual page 251

Arbitrary waveform generator
Table of Contents

Advertisement

Performance Verification
F-18
2. To set the VX4792 Arbitrary Waveform Generator controls and select the
waveform file, type:
IBWRTF GATE.CMD <RETURN>
3. To check gated mode with manual trigger:
a. Type:
IBWRT START" <RETURN>
Check that the oscilloscope continuously displays a sine wave.
b. Type:
IBWRT STOP" <RETURN>
Check that the oscilloscope continuously displays a line.
4. Check gated mode with gate signal:
a. Apply gate signal: Turn function generator output on.
b. Check gated mode with positive gate signal: Check that the oscilloscope
displays a sine wave when the gate signal is high (Figure F–7).
Ch 1 From VX4792
Ch 2 From Function
Generator
Figure F-7: Waveform Output Signal with Positive Gate Signal
c. To change the trigger polarity to negative, type:
IBWRT TRIG:POL NEG" <RETURN>
d. Check gated mode with a negative gate signal: Check that the oscillo-
scope displays a sine wave while the gate signal is low (Figure F–8).
VX4792 User Manual

Advertisement

Table of Contents
loading