Performance Verification
Check Triggered Mode
F-12
a. Connect the oscilloscope: Connect the WAVEFORM output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Set the oscilloscope controls:
Vertical
CH1 coupling
CH1 scale
CH1 input impedance:
Horizontal
Sweep
Trigger
Source
Coupling
Slope
Level
Mode
2. To set the VX4792 Arbitrary Waveform Generator controls and select the
waveform file, type:
IBWRT *RST" <RETURN>
IBWRTF SIN1000.PAT <RETURN>
IBWRTF CONT.CMD <RETURN>
3. Check against limits: Check that the oscilloscope displays a five-cycle sine
wave that has an amplitude of five divisions.
4. End procedure: Disconnect the oscilloscope.
This procedure checks the operation of the Triggered mode.
Electrical Characteristic Checked: Operating mode, Triggered, on page E–4.
Equipment Required: Two 50 W coaxial cables, a function generator, and an
oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install the test setup and set test equipment controls (see Figure F–4):
CH1
DC
0.2 V/div
50 W
5 ms/div
CH1
DC
Positive
–100 mV
Auto
VX4792 User Manual