Operating Mode Checks
Check Continuous Mode
WAVEFORM
OUTPUT
VX4792 User Manual
11. On the computer, change the directory to access IBIC.
12. To invoke IBIC and select device 1 (the VX4792 Arbitrary Waveform
Generator), type the following commands:
IBIC <RETURN>
IBFIND DEV1 <RETURN>
Now your computer will display the DEV1: prompt.
13. Perform the Functional Check Procedure located on page 1–11. This
procedure verifies that your system is operational, and runs a self-calibration
to optimize waveform generator performance.
This procedure checks the operation of the Continuous, Triggered, Burst, Gated,
Waveform Advance, and Autostep modes.
This procedure checks the operation of the Continuous mode.
Electrical Characteristic Checked: Operating mode, Continuous, on page E–4.
Equipment Required: A 50 W coaxial cable and an oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install the test setup and set test equipment controls (see Figure F–3):
VX4792
Figure F-3: Cont Mode Initial Test Setup
Performance Verification
Oscilloscope
Ch 1 Input
F-11