Table F-3: VX4792 Performance Check Disk Command File Summary (Cont.)
File Name
Contents
WADV1.CMD
*rst
mode wadv
trig:slop pos
trig:level 1v
trig:imp low
sequence:define "mode_adv.seq",#0sin1000.wfm,10
sin200.wfm,10
WADV2.CMD
wave "mode_adv.seq"
output:state on
System Setup
F-8
Before you perform the performance verification procedures, you must complete
the following tasks to configure your system. Figure F–2 shows the basic system
configuration.
1. Install the GPIB Interface circuit board and software into the IBM PC-com-
patible computer (refer to the GPIB Interface user documentation for
instructions).
2. Configure the VXIbus mainframe for the instrument modules to be installed.
See page 1–6 for Tektronix mainframes; otherwise refer to the user manual
for the mainframe for instructions.
3. Setup the Slot 0 Resource Manager for operation within this system
(Figure F–1):
H Set the logical address to 1
H Set the configuration switch for secondary addressing
(Switch #1 CLOSED)
H Set the CLK 10 switch for internal operation
H Set the interrupt handler switch to 7
Where Used