Performance Verification
Test Interval
Conventions
F-2
To ensure accurate operation of the waveform generator, check the performance
every 2000 hours, or once every 12 months if you use the waveform generator
intermittently.
Throughout the procedures within this section, the following conventions apply:
General Format. Each test procedure uses the following general format:
H Title of Test
H Equipment Required
H Prerequisites
H Procedure
Step Sequence. Each procedure consists of steps, substeps, and subparts. The
steps are arranged in the following sequence:
1. First Step.
a. First Substep.
b. Second Substep.
H First subpart.
H Second subpart.
2. Second Step.
Command Entry. When instructed to enter commands on the keyboard of your
computer, the text to be entered is formatted as shown in the following example:
1. To change the waveform generator controls, type:
IBWRT AMPL 1V" <RETURN>
where <RETURN> is a carriage return. You must type every character shown.
VX4792 User Manual