Performance Verification
F-14
d. Set the function generator controls:
Function
Mode
Parameter
Frequency
Amplitude
Offset
Output
2. To set the VX4792 Arbitrary Waveform Generator controls and select the
waveform file, type:
IBWRTF TRIG.CMD <RETURN>
NOTE. Watch the CRT of the oscilloscope when triggering manually. The
oscilloscope sweeps one time for each START command.
3. To check triggered mode with manual triggering, type:
IBWRT START" <RETURN>
Check that the oscilloscope displays a one-cycle sine wave.
4. Check triggered mode with external triggering:
a. Enable function generator output: Turn on the function generator output.
b. Check triggering: Check that for each trigger supplied by the function
generator, the oscilloscope displays a one-cycle sine wave.
5. End procedure: Turn off the function generator output, and disconnect the
function generator and oscilloscope.
Square
Continuous
100 Hz
4 V
2 V
Off