Sony Tektronix VX4792 User Manual page 155

Arbitrary waveform generator
Table of Contents

Advertisement

Command Descriptions
STARt
Group
Related Commands
Syntax
Arguments
Examples
*STB?
Group
Related Commands
Syntax
Arguments
Returns
Examples
3-72
The STARt command generates a trigger event to start the output of a waveform
or a sequence.
MODE
RUNNing?, STOP, *TRG
STARt
STARt
None
:START
generates a trigger event.
The *STB? common query returns the value of the SBR (Status Byte Register).
At this time, bit 6 of the SBR is read as a MSS (Master Status Summary) bit.
Refer to Status and Events for more details on the SBR.
STATUS and EVENT
*CLS, DESE, *ESE, *ESR, EVENT?, EVMsg?, EVQty?, *SRE
*STB?
*STB
?
None
<NR1>
which is a decimal number.
*STB?
might return 96, which indicates that the SBR contains the binary number
01100000.