Table 4-3: Clock Frequency Accuracy - Sony Tektronix AWG2021 Service Manual

Arbitrary waveform generator
Table of Contents

Advertisement

AWG2021 Service Manual
2. Set AWG2021 controls and select the waveform:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Select the waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to select the CLK_FREQ.WFM file.
H Push ENTER.
3. Check clock frequency accuracy:
a. Check clock frequency accuracy at current clock frequency setting:
Check that the frequency counter reading falls between 249.9875 MHz
and 250.0125 MHz.
b. Check clock frequency accuracy for different clock frequency settings:
H Select Clock/Divider from the bottom of the SETUP menu, or select
Clock if the instrument is not equipped with Option 02.
H Turn the general purpose knob (or press the numeric and units keys,
and push ENTER) to select the first clock frequency listed in
Table 4–3.
H Check that the frequency counter reading is within the frequency
range listed in the table for the clock frequency setting.
H Repeat this step for each clock frequency and frequency range listed
in Table 4–3.

Table 4-3: Clock Frequency Accuracy

Clock Frequency
150 MHz
1 MHz
1 kHz
10 Hz
4. End procedure: Disconnect the frequency counter.
Performance Tests
Frequency Range
149.9925 MHz - 150.0075 MHz
0.99995 MHz – 1.00005 MHz
0.99995 kHz - 1.00005 kHz
9.9995 Hz - 10.0005 Hz
4-31