Self Tests
4-4

Figure 4-1: Diagnostics Menu

b. Run the diagnostics: Select Execute from the side menu. This executes
all the AWG2021 diagnostics automatically.
c. Wait: The internal diagnostics do an extensive verification of AWG2021
functions. This verification takes about one minute. While it progresses,
the screen displays the busy icon. When finished, the resulting status
appears on the screen.
d. Confirm that no failures are found: Verify that no failures are found and
reported on-screen. If the diagnostics displays Fail as the result of any
test, use the instructions in section 6, Troubleshooting, to identify the
faulty module. If the diagnostics display an error code, contact your
nearest representative.
2. Return to regular service: Push a button (other than UTILITY) in the MENU
column to exit the diagnostic menu.