Performance Tests
Check Offset Accuracy
4-36
Electrical
Main Output, Offset, Accuracy, on page 1-9.
Characteristic
Checked
Equipment
One 50 W coaxial cable
Required
One 50 W termination
One BNC to dual banana adapter
One digital multimeter (DMM)
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Use the test hookup and test equipment settings from the previous check.
2. Check offset accuracy:
H Push the AWG2021 MANUAL TRIGGER button, and check that the
displayed step is Step 5.
H Check that the DMM voltage reading is in the range from +2.470 V to
+2.530 V.
H Push the AWG2021 MANUAL TRIGGER button, and check that the
displayed step is Step 6.
H Check that the DMM voltage reading is in the range from –0.005 V to
+0.005 V.
H Push the AWG2021 MANUAL TRIGGER button, and check that the
displayed step is Step 7.
H Check that the DMM voltage reading is in the range from –2.530 V to
–2.470 V.
3. Check Option 02: If the AWG2021 has Option 02, repeat the Gain Accuracy
and Offset Accuracy Checks for the AWG2021 channel 2 (CH2).
4. End procedure: Disconnect the DMM.