This subsection describes how to adjust internal circuits of the AWG2021.
Before doing these adjustments, read Before Adjustments on page 5–2.
Equipment
One 50 W coaxial cable
Required
One oscilloscope
Prerequisites
The AWG2021 must meet the requirements for performance listed on
page 5-1.
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure 5–2).
AWG2021
AWG2021

Figure 5-2: Hookup for X5 Amplifier HF Compensation

b. Set oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
CH1 coupling
CH1 scale
. . . . . . . . . . . . . . . . .
CH1 input impedance
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . .
Trigger
Source
. . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . .
Oscilloscope
TDS
540
CH1
. . . . . . . . . . . . . .
DC
0.1 V/div.
50 W
. . . . . . .
5 ns/div.
CH1
DC
Positive
0 V
Auto
Adjustment Procedures