External CLOCK IN Check
AWG2021 Service Manual
This procedure checks the AWG2021 response to an external CLOCK IN signal.
Electrical
Auxiliary Inputs, Clock, Threshold level, on page 1-12.
Characteristic
Checked
Equipment
Two 50 W coaxial cables
Required
One function generator
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a. Hook up oscilloscope: Connect the AWG2021 CH1 output through a
coaxial cable to the oscilloscope CH1 vertical input.
b. Hook up function generator: Connect the AWG2021 rear-panel CLOCK
IN through a coaxial cable and SMA-BNC adapter to the function
generator output (see Figure 4–18).
AWG2021
AWG2021
Figure 4-18: External CLOCK IN Initial Test Hookup
Function Generator
FG 5010
Oscilloscope
TDS
540
Performance Tests
4-45