Table 1-1: Electrical Characteristics (Cont.)
Characteristics
Main Output
Amplitude
Range
Resolution
Offset
Range
Resolution
Impedance
Auxiliary Output
SYNC
Amplitude
Impedance
MARKER1
Amplitude
Impedance
Period Jitter
Cycle to Cycle Jitter
MARKER2
Amplitude
Impedance
CLOCK
Amplitude
Impedance
AWG2021 Service Manual
Description
Except multiply (AM) and add (Add) operation
0.05 to 5 V
into 50 W
p p
The amplitude range expands about 2 times (10 V
open circuit. It's actual value is two times the displayed value.
±10 V ≥ (The absolute peak Amplitude + Offset)
1 mV
±2.5 V into 50 W
The offset range expands about 2 times (-5 V to +5 V) into
open circuit. Its actual value is two times the displayed value.
±10 V ≥ (The absolute peak Amplitude + Offset)
-100 mA to +100 mA into 50 W (current source output)
5 mV
50 W
When F.G mode and the frequency is above 250 kHz, the Sync
pulse occurs one time per two waveforms.
>1.2 V into 50 W, >2.4 V into open circuit
50 W
2.5 V (+5%, -10%) into 50 W, 5 V (+5%, -10%) into open
circuit
50 W
Refer to Table 1-1-1 (measured by TDS694C-1MHD with
TDSJIT1)
Refer to Table 1-1-2 (measured by TDS694C-1MHD with
TDSJIT1)
2.5 V (+5%, -10%) into 50 W, 5 V (+5%, -10%) into open
circuit
50 W
1 V into 50 W
50 W
Performance Characteristics
Performance Test
) into
p p
1-5